SYMPOSIUM
Year : 2012  |  Volume : 28  |  Issue : 1  |  Page : 76-81
 

Robotic laparoendoscopic single-site surgery: From present to future


1 Harran University, Department of Urology, Sanliurfa, Turkey
2 Glickman Urological and Kidney Institute, Cleveland Clinic, Cleveland, OH, USA

Date of Web Publication12-Apr-2012

Correspondence Address:
Robert J Stein
Glickman Urological and Kidney Institute, 9500 Euclid Ave, Q10, Cleveland, OH 44195
USA
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0970-1591.94962

Rights and Permissions

 
   Abstract 

The continued effort of improving cosmesis and reducing morbidity in urologic surgery has given rise to novel alternatives to traditional minimally invasive techniques: Laparoendoscopic Single-site Surgery (LESS) and Natural Orifice Transluminal Endoscopic Surgery (NOTES). Despite the development of specialized access devices and instruments, the performance of complex procedures using LESS has been challenging due to loss of triangulation and instrument clashing. A robotic interface may represent the key factor in overcoming the critical restrictions related to NOTES and LESS. Although encouraging, current clinical evidence related to R-LESS remains limited as the current da Vinci® robotic platform has not been specifically designed for LESS. Robotic innovations are imminent and are likely to govern major changes to the current landscape of scarless surgery.


Keywords: Laparoendoscopic single-site surgery, robotics, single-port laparoscopy


How to cite this article:
Verit A, Rizkala E, Autorino R, Stein RJ. Robotic laparoendoscopic single-site surgery: From present to future. Indian J Urol 2012;28:76-81

How to cite this URL:
Verit A, Rizkala E, Autorino R, Stein RJ. Robotic laparoendoscopic single-site surgery: From present to future. Indian J Urol [serial online] 2012 [cited 2019 Jul 23];28:76-81. Available from: http://www.indianjurol.com/text.asp?2012/28/1/76/94962



   Introduction Top


Urologic surgery has evolved from its origin almost 2500 years ago with the drainage of renal abscesses and removal of calculi from renal fistulas to the milestone in the last decade of the 20 th century of laparoscopic nephrectomy as described by Clayman et al.[1],[2] This new era of minimally invasive urologic surgery has gained wide acceptance clinically and has become the method of choice for many procedures. Minimally invasive surgery has resulted not only in improving cosmesis but also decreasing morbidity with decreased pain and shorter convalescence. Most recently the urologic field was introduced to robotic surgery [3] and in a relatively short period of time the great majority of prostatectomy procedures in the United States are now being performed with this technology.

The continued effort of improving cosmesis and reducing morbidity in urologic surgery has given rise to novel alternatives to traditional minimally invasive techniques: Laparoendoscopic Single-site Surgery (LESS) and Natural Orifice Transluminal Endoscopic Surgery (NOTES). Special note should be taken that in terms of nomenclature, the term LESS has been accepted as the official moniker used to refer to single-site procedures. [4],[5] In order to describe LESS procedures performed using the Da Vinci robot® (Intuitive Surgical, Sunnyvale, CA, USA) the term robotic LESS (R-LESS) has been adopted.

Robotic laparoendoscopic single-site surgery: Lights and shadows

Despite the development of specialized access devices and instruments, the performance of complex procedures using LESS has been challenging due to loss of triangulation and instrument clashing. R-LESS would seem to be uniquely suited to help overcome some of these limitations with technological advantages such as instruments with articulating wristed motion, tremor filtration, and overall ergonomic benefits that could improve surgeon comfort significantly. Furthermore, the stereoscopic three-dimensional view and the associated appreciation of the relative position of structures is a noted benefit of robotic technology. [6] The result is evidence of decreased learning curves and reduced perioperative complications for robotic compared to laparoscopic procedures. [7]

A major limitation of R-LESS is that the hardware and software of the standard robotic systems are not specifically designed for LESS and therefore clashing is still experienced and incisions may need to be somewhat larger than for conventional LESS. [8] Nevertheless, though still accounting for a small percentage of total LESS publications, there appears to be a relative increase of R-LESS procedures being performed when considering all clinical LESS cases since 2008. As will be discussed elsewhere in this review, R-LESS may be controversial since most procedures are extirpative or ablative rather than reconstructive and commonly lack ancillary procedures such as lymph node dissection during radical prostatectomy and hilar clamping during partial nephrectomy. [9],[10]

Access techniques

In terms of access, transvaginal, transrectal and transumbilical routes have been used either clinically or experimentally for R-LESS as well as hybrid NOTES approaches. [10] The present access technique for upper tract R-LESS involves a peri-umbilical "omega" Ω-like skin incision in a range of 4-5 cm after placing the patient in 60-90° flank position with a slight break in the table. For pelvic R-LESS procedures the patient is placed in the Trendelenburg dorsal lithotomy position and an "omega"- like or midline umbilical incision is created. [11] Due to pneumoperitoneum leakage commonly experienced with robotic trocars placed through a single-port device, it is suggested to place the robotic instrument trocars via separate fascial stabs alongside the device or use a single-incision, multiport technique without a commercially available device. The authors prefer a 30° downward viewing robotic scope for upper tract procedures and a 0° scope for pelvic procedures. Five and 8-mm robotic ports may be used but the authors prefer using 8-mm instruments for both robotic arms. Thus far, the use of a fourth arm has not been described. Short length of the instruments may be another issue theoretically as the ports are positioned caudally, especially for dissection of the upper pole of the kidney. As the da Vinci S and Si systems® offer 5-cm longer robotic arms compared to the standard version the authors have not noted difficulty in terms of reach. [12] Of note, as with all LESS procedures patient selection is important and the body mass index (BMI) of initial reported R-LESS patients was in the range of 23-30 kg/m 2 . [13]

Robotic NOTES

Potential advantages for robotic NOTES (R-NOTES) include even better cosmesis, improved access in obese patients or for those with abdominal wall pathologies, less pain, and avoidance of abdominal wall trauma. Although to date there has been no clinical report of R-NOTES in urologic surgery, Kaouk et al., reported a pure NOTES transvaginal nephrectomy with articulating and conventional laparoscopic instruments. [14] All in all, unlike R-LESS which appears to be gaining some clinical momentum, R-NOTES is still largely in a developmental stage.

Access platforms

A wide variety of access platforms from homemade to commercially available have been used in R-LESS. The most common are, i) Quadport® (Advanced Surgical Concepts, Wicklow, Ireland) which contains 5(1), 10(2) and 15(1)-mm ports and is placed through a 2-7 cm incision, [15] ii) SILS® port (Covidien, Mansfield, MA, USA) which contains four sites for port placement and is introduced through a 3-4cm incision, iii) GelPort/GelPoint® (Applied Medical, Rancho Santa Margarita, California, USA) which include a GelCap through which various ports in any arrangement may be introduced. [16] The optimal incision for placement of these ports is "3-7 cm, and iv)" Homemade access platforms composed of a wound retractor as an inner ring covered by an intact surgical glove with trocars introduced through the fingers of the glove. [17],[18] Besides these, a novel robotic device has recently been approved for clinical use in Europe and is used with a multichannel port designed for placement of an 8.5-mm scope, an assistant 12-mm cannula and two additional curved cannulas for the robotic arms [8] [Figure 1]. For all commercially available ports, a disadvantage includes the cost of the device. [19] Presently, the majority of devices average approximately $400 in the United States with differences depending upon institutional negotiation. In the author's experience, when using the SILS port, placement of the robotic trocars through the same incision but through separate fascial stabs alongside the port helps with increased instrument spacing and decreased leakage of insufflant. Also, use of multiple ports introduced through separate fascial stabs using a single incision can be successful and is more cost-effective [Figure 1].
Figure 1: Port positioning and docking of the robot during R-LESS prostatectomy without a commercially available single-port device

Click here to view


Initial experimental studies

Initial experimental studies have been conducted on pigs and human cadavers. Box et al., performed R-NOTES nephrectomy on a female farm pig with robotic arms docked through the umbilicus, vagina and rectum. However, they reported that this atypical trocar configuration led to difficulty in instrument movement despite being able to supply effective traction for exposure and dissection. [12] Haber et al., reported performing 30 urologic R-LESS procedures in 10 female farm pigs with transumbilical access for the camera and one robotic arm and transvaginal access for the other robotic arm. [7] Procedures performed included dismembered pyeloplasty, partial nephrectomy, and radical nephrectomy. The authors concluded that the robot enhanced intracorporeal suturing but that there remained significant clashing of the arms despite greater separation by using the vagina as an access site. They also noted that R-NOTES performed on pigs was likely technically easier than on humans due to the anatomy of the porcine model such as minimal Gerota's fat, smaller renal vessels and shorter distance between the kidney and vaginal access for upper tract R-NOTES/LESS. Haber et al. , in a more recent report explored the feasibility and efficiency of a novel modification of the da Vinci Si surgical system for R-LESS (VeSPA) with a single-port device and curved cannulas [Figure 2]. They performed 16 procedures including radical nephrectomy (n=8), partial nephrectomy (n=4), and dismembered pyeloplasty (n=4) and reported that while no extracorporeal clashing between the robotic arms occurred, the articulating function of the endo-wrists was not included on the present prototype. [18]
Figure 2: (a) Curved trocars, fl exible instruments, and specially designed port of the VeSPA robotic system, (b) Patient cart docked to the VeSPA robotic system

Click here to view


R-LESS was used to perform transvesical robotic radical prostatectomy in two human cadavers by either a multiport device with four separate ports introduced percutaneously into the bladder or by a Quadport® device. After radical excision of the prostate, urethrovesical anastamosis could be completed but with some difficulty. [20] The authors encourage future attempts using the transvesical route for R-LESS based on experience of a large intravesical working space upon establishment of pneumovesicum. However, inability to perform lymph node dissection indicated for moderate and high-risk prostate cancer cases was a limitation for this novel method.

Human experience

Based on these initial experimental studies, Kaouk et al., published their initial R-LESS case series [Table 1] involving upper and lower tract common urologic extirpative and reconstructive procedures including radical nephrectomy, pyeloplasty and radical prostatectomy. [21] Procedures were performed through a transumbilical incision with the insertion of a robotic 12-mm scope and a 5-mm instrument port via the R-Port system. Additionally, a 5 or 8-mm robotic port was introduced through a separate fascial stab within the same skin incision. The authors suggested that the learning curve for R-LESS would likely be shorter in comparison with standard LESS. Furthermore, the same center soon after published their experience of R-LESS partial nephrectomy for two patients. [21] The authors introduced a Triport (Advanced Surgical Concepts, Wicklow, Ireland) through a 1.8-cm umbilical incision. The platform provided a 10-mm port for introduction of the robotic camera and an additional 5-mm robotic instrument was introduced through the device. A separate robotic port was placed through a separate fascial stab alongside the device and through the same incision. The procedures were done without hilar clamping and the procedure was suggested only for exophytic and middle and lower pole tumors. To expand the indications for partial nephrectomy, Arkoncel et al., suggested a two-incision hybrid R-LESS partial nephrectomy. [22],[23] The procedure was used to perform 35 partial nephrectomies and the additional 12-mm port was placed 8 cm caudal to the homemade umbilical access device and used for assistance and placement of hilar clamps. Another clinical series described the use of the GelPort platform for R-LESS in 11 patients undergoing robotic prostatectomy, partial/radical nephrectomy, ureteroneocystostomy [Figure 3] and dismembered pyeloplasty. [16] The GelPort was noted to facilitate the procedures by providing greater spacing of ports and greater flexibility in terms of optimal trocar positioning. Barret et al. , described their initial experience with extraperitoneal R-LESS radical prostatectomy in a cadaver followed by a clinical case. No commercial access device was used as they placed all robotic ports through separate fascial punctures transumbilically. An additional 5-mm port through a separate skin incision was used by the bedside assistant. [24] R-LESS prostatectomy has also been described for a patient who underwent neoadjuvant hormonal deprivation therapy. [25] Seo et al., reported a hybrid R-LESS approach for simultaneous bilateral upper tract procedures. [26] They performed partial nephrectomy for a 51-year-old man (BMI 23 kg/m 2 ) for bilateral renal masses with a 3-cm umbilical incision using a homemade access device. An additional 11-mm trocar was added for the bedside assistant. While one of the partial nephrectomies required renal vascular clamping, the other did not. External or internal clashing was the most common difficulty for the surgeons with R-LESS. Joseph et al. , described a 'chopstick' technique used in the laboratory in order to reduce instrument clashing. The technique involves crossing of the robotic arms inside the patient and modifying the outputs of the robot such that the 'left' instrument is controlled with the right hand effector and vice versa. [27] In our experience this technique can be useful but oftentimes cannot be used for entire procedures.
Figure 3: (a) Illustration of port positioning for R-LESS ureteroneocystostomy. Note the placement of one robotic trocar beside the Quadport through the same incision but a separate fascial stab, (b) Illustration of R-LESS ureteroneocystostomy using two 5-mm robotic needle drivers

Click here to view
Table 1: Robotic urologic LESS: reported clinical series*

Click here to view


The future of NOTES and LESS

R-LESS indeed offers many potential advantages over standard LESS such as superior ergonomics and easier instrument articulation, yet there are still significant limitations including the need for a somewhat larger incision with the present robot, clashing, lack of haptic feedback, and limited space for the assistant to maneuver. The advantages of LESS for urologic procedures are still undefined and somewhat controversial. [28] What is clear is that LESS is being performed increasingly as the techniques are refined.

NOTES and LESS, in their robotic form, are rapidly developing new techniques that inevitably depend on cooperative support from engineers, clinical researchers and minimally invasive surgeons. This enormous, multidisciplinary effort will be more likely directed towards the development of more sophisticated robotic devices, such as flexible robots and in vivo wireless controlled miniature robots. [29] Several flexible robotic systems have been already introduced primarily for intravascular catheter-based applications. Multiple remotely controlled surgical instruments and scopes through a common channel can give the opportunity to the surgeon to perform complex intra-abdominal procedures by providing the necessary triangulation, degrees of freedom, and tensile strength. The working channels of this robotic instrument could carry computer-controlled flexible graspers, needle drivers, mono-bipolar coagulators and even ultrasonic scalpels and suction arms. The initial version of this flexible robotic system has been used experimentally for ureterorenoscopy. [30],[31] Also, in-vivo mini robots have been introduced into the peritoneal cavity experimentally via transgastric incision to perform NOTES small bowel dissection. [32],[33]

Magnetic Anchoring and Guidance system (MAGS) is another interesting technology in the development of LESS. Basically, the system is composed of intracorporeal LESS instruments anchored through the abdominal wall with extracorporeal magnetic devices. This system was successfully tested in performing a LESS nephrectomy. [34] It was claimed that the MAGS camera results in fewer instrument collisions and allows for an adequate working space despite its current obstacles such as a fixed zero-degree lens and focus, as well as magnets that require a thin abdominal wall. [35]

Newer robotic surgical systems are on the horizon. The ViKY® system (EndoControl, Grenoble, France) with a compact camera holder will have the ability of supplying a 5-mm endoscope without decreasing viewing capacity, and 3-mm trocars. [6],[31] The SPIDER® surgical system (TransEnterix, Durham, North Carolina, USA) is designed to be introduced through a small incision but has curved trocars through which flexible instruments can be introduced. It has been recently experimentally and clinically tested by means of performing simple LESS cases and the authors were impressed by its performance due to its superior triangulation without clashing. However, the authors had great difficulty in basic maneuvers such as dissecting, retracting, grasping, cutting and cauterizing. The second-generation SPIDER system evolved to a vertebral design in order to increase the forces generated at the distal instrument tips and thereby overcome the tensile strength limitations. [36]


   Conclusions Top


A robotic interface may represent the key factor in overcoming the critical restrictions related to NOTES and LESS. Although encouraging, current clinical evidence related to R-LESS remains limited as the current da Vinci® robotic platform has not been specifically designed for LESS. Robotic innovations are imminent and are likely to govern major changes to the current landscape of scarless surgery.

 
   References Top

1.Lima E, Rolanda C, Pêgo JM, Henriques-Coelho T, Silva D, Osório L, et al. Third-generation nephrectomy by natural orifice transluminal endoscopic surgery. J Urol 2007;178:2648-54.  Back to cited text no. 1
    
2.Clayman RV, Kavoussi LR, Soper NJ, Dierks SM, Meretyk S, Darcy MD, et al. Laparoscopic nephrectomy: Initial case report. J Urol 1991;146:278-82.  Back to cited text no. 2
    
3.Menon M, Shrivastava A, Tewari A, Sarle R, Hemal A, Peabody JO, et al. Laparoscopic and robot assisted radical prostatectomy: Establishment of a structured program and preliminary analysis of outcomes. J Urol 2002;168:945-9.  Back to cited text no. 3
    
4.Box G, Averch T, Cadeddu J, Cherullo E, Clayman R, Desai M, et al; Urologic NOTES Working Group. Nomenclature of natural orifice translumenal endoscopic surgery (NOTES) and laparoendoscopic single-site surgery (LESS) procedures in urology. J Endourol 2008;22:2575-81.  Back to cited text no. 4
    
5.Gettman MT, Box G, Averch T, Cadeddu JA, Cherullo E, Clayman RV, et al. Consensus statement on natural orifice transluminal endoscopic surgery and single-incision laparoscopic surgery: Heralding a new era in urology? Eur Urol 2008;53:1117-20.  Back to cited text no. 5
    
6.Micali S, Pini G, Teber D, Sighinolfi MC, De Stefani S, Bianchi G, et al. New trends in minimally invasive urological surgery. What is beyond the robot? World J Urol 2010 Sep 2 [DOI: 10.10071s00345-010-0588-5].   Back to cited text no. 6
    
7.Patel HR, Linares A, Joseph JV. Robotic and laparoscopic surgery: Cost and training. Surg Oncol 2009;18:242-6.  Back to cited text no. 7
    
8.Haber GP, White MA, Autorino R, Escobar PF, Kroh MD, Chalikonda S, et al. Novel robotic daVinci instruments for laparoendoscopic single-site surgery. Urology 2010;76:1279-82.  Back to cited text no. 8
    
9.Kaouk JH, Autorino R, Kim FJ, Han DH, Lee SW, Yinghao S, et al. Laparoendoscopic Single-site Surgery in Urology: Worldwide Multi-institutional Analysis of 1076 Cases. Eur Urol 2011;60:998-1005.  Back to cited text no. 9
    
10.Autorino R, Stein RJ, Lima E, Damiano R, Khanna R, Haber GP, et al. Current status and future perspectives in laparoendoscopic single-site and natural orifice transluminal endoscopic urological surgery. Int J Urol 2010;17:410-31.  Back to cited text no. 10
    
11.White MA, Haber GP, Autorino R, Khanna R, Altunrende F, Yang B, et al. Roboticlaparoendoscopic single-site surgery. BJU Int 2010;10:923-7.  Back to cited text no. 11
    
12.Box GN, Lee HJ, Santos RJ, Abraham JB, Louie MK, Gamboa AJ, et al. Rapid communication: Robot-assisted NOTES nephrectomy: Initial report. J Endourol 2008;22:503-6.  Back to cited text no. 12
    
13.Kaouk JH, Goel RK, Haber GP, Crouzet S, Stein RJ. Robotic single-port transumbilical surgery in humans: Initial report. BJU Int 2009;103:366-9.   Back to cited text no. 13
    
14.Kaouk JH, Haber GP, Goel RK, Crouzet S, Brethauer S, Firoozi F, et al. Pure natural orifice translumenal endoscopic surgery (NOTES) transvaginal nephrectomy. Eur Urol 2010;57:723-6.   Back to cited text no. 14
    
15.Granberg CF, Gettman MT. Instrumentation for natural orifice translumenal endoscopic surgery and laparoendoscopic single-site surgery. Indian J Urol 2010;26:385-8.  Back to cited text no. 15
[PUBMED]  Medknow Journal  
16.Stein RJ, White WM, Goel RK, Irwin BH, Haber GP, Kaouk JH. Robotic laparoendoscopic single-site surgery using GelPort as the access platform. Eur Urol 2010;57:132-6.   Back to cited text no. 16
    
17.Jeon HG, Jeong W, Oh CK, Lorenzo EI, Ham WS, Rha KH, et al. Initial experience with 50 laparoendoscopic single site surgeries using a homemade, single port device at a single center. J Urol 2010;183:1866-71.   Back to cited text no. 17
    
18.Han WK, Kim DS, Jeon HG, Jeong W, Oh CK, Choi KH, et al. Robot-assisted laparoendoscopic single-site surgery: Partial nephrectomy for renal malignancy. Urology 2010;77:612-6.   Back to cited text no. 18
    
19.Dev H, Sooriakumaran P, Tewari A, Rane A. LESSons in minimally invasive urology. BJU Int 2011;107:1555-9.  Back to cited text no. 19
    
20.Desai MM, Aron M, Berger A, Canes D, Stein R, Haber GP, et al. Transvesical robotic radical prostatectomy. BJU Int 2008;102:1666-9.   Back to cited text no. 20
    
21.Kaouk JH, Goel RK. Single-port laparoscopic and robotic partial nephrectomy. Eur Urol 2009;55:1163-9.   Back to cited text no. 21
    
22.Han WK, Kim DS, Jeon HG, Jeong W, Oh CK, Choi KH, et al. Robot-assisted laparoendoscopic single-site surgery: Partial nephrectomy for renal malignancy. Urology 2011;77:612-6.  Back to cited text no. 22
    
23.Arkoncel FR, Lee JW, Rha KH, Han WK, Jeoung HB, Oh CK. Two-port Robot-assisted vs Standard Robot-assisted Laparoscopic Partial Nephrectomy: A Matched-pair Comparison. Urology 2011;78:581-5.  Back to cited text no. 23
    
24.Barret E, Sanchez-Salas R, Kasraeian A, Benoist N, Ganatra A, Cathelineau X, et al. A transition to laparoendoscopic single-site surgery (LESS) radical prostatectomy: Human cadaver experimental and initial clinical experience. J Endourol 2009;23:135-40.  Back to cited text no. 24
    
25.Leewansangtong S, Vorrakitkatorn P, Amornvesukit T, Taweemonkongsap T, Nualyong C, Sujijantararat P. Laparo-endoscopic single site (LESS) robotic radical prostatectomy in an Asian man with prostate cancer: An initial case report. J Med Assoc Thai 2010;93:383-7.  Back to cited text no. 25
    
26.Seo IY, Rim JS. Bilateral robotic single-site partial nephrectomy. J Laparoendosc Adv Surg Tech A 2011;21:435-8.  Back to cited text no. 26
    
27.Joseph RA, Goh AC, Cuevas SP, Donovan MA, Kauffman MG, Salas NA, et al. "Chopstick" surgery: A novel technique improves surgeon performance and eliminates arm collision in robotic single-incision laparoscopic surgery. Surg Endosc 2010;24:1331-5.   Back to cited text no. 27
    
28.Gettman MT, White WM, Aron M, Autorino R, Averch T, Box G, et al; Endourological Society NOTES and LESS Working Group; European Society of Urotechnology NOTES and LESS Working Group. Where do we really stand with LESS and NOTES? Eur Urol 2011;59:231-4.  Back to cited text no. 28
    
29.Canes D, Lehman AC, Farritor SM, Oleynikov D, Desai MM. The future of NOTES instrumentation: Flexible robotics and in vivo minirobots. J Endourol 2009;23:787-92.   Back to cited text no. 29
    
30.Aron M, Desai MM. Flexible robotics. Urol Clin North Am 2009;36:157-62.  Back to cited text no. 30
    
31.Teber D, Baumhauer M, Guven EO, Rassweiler J. Robotic and imaging in urological surgery. Curr Opin Urol 2009;19:108-13.  Back to cited text no. 31
    
32.Lehman AC, Dumpert J, Wood NA, Visty AQ, Farritor SM, Varnell B, et al. Natural Orifice Translumenal Endoscopic Surgery with a miniature in vivo surgical robot. Surg Endosc 2009;23:1649.  Back to cited text no. 32
    
33.Dumpert J, Lehman AC, Wood NA, Oleynikov Farritor SM. Semi-autonomous surgical tasks using a miniature in vivo surgical robot. Conf Proc IEEE Eng Med Biol Soc 2009;2009:266-9.  Back to cited text no. 33
    
34.Cadeddu J, Fernandez R, Desai M, Bergs R, Tracy C, Tang SJ, et al. Novel magnetically guided intra-abdominal camera to facilitate laparoendoscopic single-site surgery: Initial human experience. Surg Endosc 2009;23:1894-9.  Back to cited text no. 34
    
35.Best SL, Cadeddu JA. Development of magnetic anchoring and guidance systems for minimally invasive surgery. Indian J Urol 2010;26:418-22.  Back to cited text no. 35
[PUBMED]  Medknow Journal  
36.Haber GP, Autorino R, Laydner H, Yang B, White MA, Hillyer S, et al. SPIDER surgical system for urologic procedures with laparoendoscopic single-site surgery: From initial laboratory experience to first clinical application. Eur Urol 2012 Feb; 61:415-22  Back to cited text no. 36
    


    Figures

  [Figure 1], [Figure 2], [Figure 3]
 
 
    Tables

  [Table 1]

This article has been cited by
1 Laparoscopy-assisted Total Gastrectomy: A Simplified Approach
Kazuya Muguruma,Hiroaki Tanaka,Katsunobu Sakurai,Takahiro Toyokawa,Naoshi Kubo,Yoshito Yamashita,Tetsuji Sawada,Masaichi Ohira,Kosei Hirakawa
International Surgery. 2014; 99(1): 79
[Pubmed] | [DOI]
2 Achieving scarless renal surgery: is it feasible?
Li June Tay,Marios Hadjipavlou,Shahid Khan,Abhay Rane
Clinical Practice. 2014; 11(6): 711
[Pubmed] | [DOI]
3 Robotic single-site hysterectomy: feasibility, learning curve and surgical outcome
V. Cela,L. Freschi,G. Simi,M. Ruggiero,R. Tana,N. Pluchino
Surgical Endoscopy. 2013; 27(7): 2638
[Pubmed] | [DOI]



 

Top
Print this article  Email this article
 

    

 
   Search
 
  
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Article in PDF (1,169 KB)
    Citation Manager
    Access Statistics
    Reader Comments
    Email Alert *
    Add to My List *
* Registration required (free)  


    Abstract
   Introduction
   Conclusions
    References
    Article Figures
    Article Tables

 Article Access Statistics
    Viewed3374    
    Printed107    
    Emailed1    
    PDF Downloaded113    
    Comments [Add]    
    Cited by others 3    

Recommend this journal

HEALTHWARE INDIA